Instrument Disinfectant

REPRODIS HLD4l Instrument Disinfectant.

Application: 

Use of REPRODIS in Machines and tanks

  • After using REPROZYME MANUAL to clean instruments, make up a solution of 1 in 20 REPRODIS HLD4l into an open tank or ultrasonic bath.
  • Remove instruments from REPROZYME solution and rinse in water of suitable quality.
  • Completely immerse the instruments into the freshly made up solution of REPRODIS HLD4l for 10 minutes.
  • Remove the instruments after the 10 minute contact time and rinse in fresh water.
  • Dry instruments and store them correctly ready for reuse.
  • The solution can last up to 7 days, provided the solution does not become contaminated or cloudy.

View Catalog

REPROZYME Enzymatic Instrument Cleaner.

Application: 

For use in Ultrasonic chambers:

  • Rinse instruments under running water.
  • Shake off residual water.
  • Make up a 2% solution of REPROZYME and immerse instruments.
  • Soak for 5 minutes.
  • Do not exceed temperature of 40°C
  • Remove, rinse and dry prior to autoclave or chemical disinfection with REPRODIS.

For use as a soak solution in non-ultrasonic tank:

  • Follow directions for ultrasonic use up to and including preparation of 2% solution.
  • Soak for 10 minutes.
  • Remove, rinse and dry prior to autoclave or chemical disinfection with REPRODIS.

View Catalog